The multi-detector optical configuration assures long-term stability and minimizes stray light and color interference. The new ratiometric design allows for easy and accurate testing. The nephelometric mode measures 0-40 NTU/FNU, ratiometric mode 40-1000 NTRU/FNRU and 1000-2000 AU. Now pinpoint the range of interest with better, more reliable results. Ideally suited for low-level drinking water applications, mid-range industrial applications, and high-range environmental applications.
version meets US EPA design criteria as specified by EPA 180.1, Rev. 2.0 (1993) and Standard Methods 2130 B-2001.
version meets design criteria for quantitative methods of turbidity using optical turbidimeters as specified by ISO 7027. Best for colored or extremely turbid samples.
Kits are supplied with 0, 1, and 10 NTU standards (Code 1974-T) or 0, 10, 100 FNU (Code 1974-I)
🟢 This product has no import restrictions.
Mode |
Ratiometric |
Nephelometric |
Attenuation |
---|---|---|---|
Unit of Measure 2020t: | NTRU, NTU, ASBC, EBC | NTU, ASBC, EBC | AU, NTU, ASBC, EBC |
Unit of Measure 2020I: | FNRU, NTU, ASBC, EBC | FNU, NTU, ASBC, EBC | FAU, NTU, ASBC, EBC |
Range: | 0-1,000 NTRU/FNRU 0-17,500 ASBC 0-250 EBC |
0-100 NTU/FNU 0-1,750 ASBC 0-25 EBC |
0-2,000 AU/FAU 0-35,000 ASBC, 0-500 EBC |
Resolution: | 0-10.99 NTRU/FNRU: 0.01 11.0-109.9 NTRU/FNRU: 0.1 110-1000 NTRU/FNRU: 1 |
0-10.99 NTU/FNU: 0.01 11.0-100.0 NTU/FNU: 0.1 |
0-2000 AU/FAU: 1 |
Accuracy: | 0-2.5 NTRU/FNRU: ±0.05 2.5-100 NTRU/FNRU: ±2% 100-1000 NTRU/FNRU: ±3% |
0-2.5 NTU/FNU: ±0.05 2.5-100 NTU/FNU: ±2% |
0-2000 AU/FAU: ±10 AU/FAU or 6%, whichever is greater |
Detection Limit: | 0.05 NTRU/FNRU | 0.05 NTU/FNU | 10 AU/FAU |
Range Selection: | Automatic | ||
Light Source: | 2020t: Tungsten lamp 2300° K ±50° K; | 2020i: IR LED 860 nm ±10 nm, spectral bandwidth with 50 nm | |
Detector: | 2020t: Photodiode, centered at 90° and 180°, maximum peak 400-600 nm; | 2020i: Photodiode, centered at 90° and 180° | |
Meter Specifications | |||
Signal Averaging: | Disabled, 2, 5, 10 | ||
Power: | USB computer cable, wall adapter or Lithium ion rechargeable battery, 3.7V, 2.5" x .75", 1.7 oz | ||
Data Logging: | 500 points | ||
Auto Shut-Off: | Disabled, 5, 10, 30 seconds | ||
Languages: | English, French, Spanish, Japanese, Italian, Portuguese, Chinese | ||
Response Time: | 2 seconds | ||
Size: | 7.5 x 3.5 x 2.5 inches | ||
Weight: | 13 ounces | ||
Display: | 6-line LCD with backlit display |